Industry
Healthcare · Medical devices
Client
Löwenstein Medical
Tools
Sketching, SolidWorks, Cinema 4D with Corona Renderer
Ventilation Masks
Medical device for non-invasive ventilation

A full-face ventilation mask that stays sealed under high pressure and lets patients sleep.
About
Patients with respiratory insufficiency wear a ventilation mask every night, often for years. A mask that leaks, presses or makes noise wakes them, and can make them stop therapy.
Worked on
· Form language · Materials and handling details · Ergonomics & usability · Sketching & CAD · Prototyping · 3D visualization
Achieved
Löwenstein's engineering and clinical teams set requirements, key dimensions and therapy targets. We turned a functional model into a mask with three cushion sizes on one body, fast emergency removal and simple disassembly.
customer website

Challenge
A working functional model, but not usable or designed.
Löwenstein's engineers brought a working functional model built to fixed requirements. It had no form language, no intuitive handling and no clear way to take it apart for cleaning.
How it came together
Key steps
01
Requirements to form
Engineering and clinical teams set dimensions, materials and therapy targets. I turned them into a usable, comfortable, manufacturable form through sketches and SolidWorks studies.

02
Interchangeable parts
I modelled the removable parts and the connecting details in the mask body, form and mechanism. Cushions swapped in, the mask came apart, and the shapes guided the hand.
03
Prototyping
We tested physical prototypes for facial fit, assembly logic and stability. The findings reshaped the cushion geometry and frame proportions: more comfort, same seal.
04
Production & patents
With engineering I detailed the moulded frame, textile headgear, silicone cushion and O₂ and exhalation integration. We mapped patents early and changed mechanisms where needed.
05
3D visualisation

Learnings from this project
Ergonomics is a balance, not an optimum
Seal, pressure, movement and comfort pulled in different directions. The design had to hold up across all of them.
Constraints are design inputs
Safety, cleaning, durability, risk, materials and patents shaped the design from day one, not as a final check.
Test early
Prototypes exposed sealing, movement and handling problems that CAD couldn’t predict.
Technology & behaviour
No screen, so the form does the talking
Without electronics, every signal came from the product itself: grips showed where to pull, the release could be found by touch in the dark, and diffuse exhalation kept the mask quiet enough to sleep.
Double-lip silicone cushion
Seal that holds under high, fluctuating BiLevel pressure
Diffuse exhalation
Quiet breathing out, no noise that wakes the patient
Ball-and-socket hose joint
Mask stays in place when the patient turns
Integrated O₂ port
Oxygen added without extra parts on the face
Vented and non-vented versions
One design for home therapy and clinical use, including endoscopy