Industry

Healthcare · Medical devices

Client

Löwenstein Medical

Tools

Sketching, SolidWorks, Cinema 4D with Corona Renderer

Ventilation Masks

Medical device for non-invasive ventilation

Main Project Image

A full-face ventilation mask that stays sealed under high pressure and lets patients sleep.

About

Patients with respiratory insufficiency wear a ventilation mask every night, often for years. A mask that leaks, presses or makes noise wakes them, and can make them stop therapy.

Worked on

· Form language · Materials and handling details · Ergonomics & usability · Sketching & CAD · Prototyping · 3D visualization

Achieved

Löwenstein's engineering and clinical teams set requirements, key dimensions and therapy targets. We turned a functional model into a mask with three cushion sizes on one body, fast emergency removal and simple disassembly.

customer website

Challenge

A working functional model, but not usable or designed.

Löwenstein's engineers brought a working functional model built to fixed requirements. It had no form language, no intuitive handling and no clear way to take it apart for cleaning.

How it came together

Key steps

01

Requirements to form

Engineering and clinical teams set dimensions, materials and therapy targets. I turned them into a usable, comfortable, manufacturable form through sketches and SolidWorks studies.

02

Interchangeable parts

I modelled the removable parts and the connecting details in the mask body, form and mechanism. Cushions swapped in, the mask came apart, and the shapes guided the hand.

03

Prototyping

We tested physical prototypes for facial fit, assembly logic and stability. The findings reshaped the cushion geometry and frame proportions: more comfort, same seal.

04

Production & patents

With engineering I detailed the moulded frame, textile headgear, silicone cushion and O₂ and exhalation integration. We mapped patents early and changed mechanisms where needed.

05

3D visualisation

Learnings from this project

Ergonomics is a balance, not an optimum

Seal, pressure, movement and comfort pulled in different directions. The design had to hold up across all of them.

Constraints are design inputs

Safety, cleaning, durability, risk, materials and patents shaped the design from day one, not as a final check.

Test early

Prototypes exposed sealing, movement and handling problems that CAD couldn’t predict.

Technology & behaviour

No screen, so the form does the talking

Without electronics, every signal came from the product itself: grips showed where to pull, the release could be found by touch in the dark, and diffuse exhalation kept the mask quiet enough to sleep.

Double-lip silicone cushion

Seal that holds under high, fluctuating BiLevel pressure

Diffuse exhalation

Quiet breathing out, no noise that wakes the patient

Ball-and-socket hose joint

Mask stays in place when the patient turns

Integrated O₂ port

Oxygen added without extra parts on the face

Vented and non-vented versions

One design for home therapy and clinical use, including endoscopy

Let’sprototypetogether!